READ POSITION Command (34h) SDLT 600 SCSI Interface Guide
4-192 January 2005 81-81200-02 REV A01
4.18.2 Read Position Data — Long Form
Data returned when the service action value is 6 takes the following form:
Figure 4-101. READ POSITION (Long Form) — Data Format
Bit
Byte
7 6 5 4 3 2 1 0
0 BOP EOP Reserved MPU LONU Reserved
1 – 3 Reserved
4 – 7
(MSB)
Partition Number
(LSB)
8 – 15
(MSB)
Logical Object Number
(LSB)
16 – 23
(MSB)
Logical File Identifier
(LSB)
24 – 31
(MSB)
Logical Set Identifier
(LSB)
Table 4-112. READ POSITION (Long Form) Data — Field Descriptions
Field Name Description
BOP Beginning of Partition. When set to 1, indicates that the drive is at the
Beginning of Partition (BOP) in the current partition. When set to 0,
indicates that the current logical position is not at BOP. Since SDLT tape
drives do not support more than one partition, the value of this field is 1
when at BOT.
EOP End of Partition. When set to 1, indicates that the drive is positioned
between early warning and the End of Partition (EOP) in the current
partition. When set to 0, indicates that the current logical position is not
between early warning and EOP.